Tube 66S
1. PERFORMANCE
1) Measuring range :1-50 ppm
Number of pump strokes :1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :0.2 ppm
4) Shelf life :1 year (Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Bromine is produced by an Oxidizer. By reacting between this Bromine and o-Toluidine, Orthoquinone is produced
BrCH2CH2Br+I(xiàn)2O4+CrO3+H2SO4→Br2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
DETECTOR TUBE
(Olive) (White)
(Orange)
PRETREAT TUBE
PRETREAT REAGENT
50
40
30
30
10
5
1
ppm
100ml
166S ETHYLENE DIBROMIDE
Substance Interference ppm Coexistence
Halogens Similar stain is produced. Higher readings are given.
Halogenated hydrocarbons 〃〃
Hexane FIG.1 aTfhfeec atecdc.uracy of readings is not 200 Lower readings are given.
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 10℃ 15℃ 20℃ 25℃ 30℃ 35℃
(ppm) (50°F) (59°F) (68°F) (77°F) (80°F) (95°F)
50 - - 82 50 42 39
40 - 80 58 40 35 33
30 98 56 40 30 27 26
20 50 40 30 20 18 18
10 16 14 12 10 10 10
5 7 7 6 5 5 5
1 1 1 1 1 1 1
0
50
60
70
80
40
30
20
10
Hexane〔ppm〕?
FIG.1 Influence of Hexane
Ethylene Dibromide indicating value 〔ppm〕
100 200 300 400 500 600 700

| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |